Skip to main content
Waters

Source pressure test failure after cleaning the cone - WKB29091

Article number: 29091

SYMPTOMS

  • Source pressure test failure
  • Analyzer vacuum failure
  • Found: system is vented
  • Found: cone maintenance performed; cone reinstalled shortly before vacuum failure

ENVIRONMENT

  • Xevo TQ-XS

CAUSE

Sample cone is not seated properly.

FIX or WORKAROUND

  1. Close the isolation valve.
  2. Start the vacuum system.
  3. Allow to pump for 1 hour with the isolation valve closed.
  4. Open the isolation valve.

ADDITIONAL INFORMATION

When the isolation valve is open, the vacuum draw will pull the cone into a better-seated position.

Do not leave the instrument in Operate with the isolation valve closed.

Not able to find a solution? Click here to request help.