Source pressure test failure after cleaning the cone - WKB29091
Article number: 29091
SYMPTOMS
- Source pressure test failure
- Analyzer vacuum failure
- Found: system is vented
- Found: cone maintenance performed; cone reinstalled shortly before vacuum failure
ENVIRONMENT
- Xevo TQ-XS
CAUSE
Sample cone is not seated properly.
FIX or WORKAROUND
- Close the isolation valve.
- Start the vacuum system.
- Allow to pump for 1 hour with the isolation valve closed.
- Open the isolation valve.
ADDITIONAL INFORMATION
When the isolation valve is open, the vacuum draw will pull the cone into a better-seated position.
Do not leave the instrument in Operate with the isolation valve closed.