Error using OpenLynx to process data via sample list; processing halts and OpenLynx Method Editor opens - WKB26491
Article number: 26491
SYMPTOMS
- Processing data file in sample list using OpenLynx
- Processing halts with "Sample 1 Processing" in the status bar above the sample list
- "Not responding" appears at the top of the MassLynx window
- Error messages may appear:
- "C:\Masslynx\<OpenLynx method name> was not found"
- "This file is an old version. The integration parameters must be re-entered"
- The OpenLynx Editor's Setup window opens
- The .rpt file is not generated
ENVIRONMENT
- MassLynx
- OpenLynx
CAUSE
Corrupted sample list, possibly caused by an invisible entry in a hidden/not displayed column cell
FIX or WORKAROUND
- Create a new sample list based on the default sample list in the Default project.
- Remove the faulty sample list from the active project (may require extra permissions).