How do you configure an inlet system in MassLynx? - WKB263023
Article number: 263023
ENVIRONMENT
- MassLynx
ANSWER
- Open the MassLynx Inlet Method Editor.
- Go to Tools > Instrument Configuration.
- Hit the Configure button.
- Follow the Wizard, selecting the appropriate Pump (Solvent Manager) and Autosampler (Sample Manager).
- On the detector page, select up to four detectors from the list on the left and add them to the list on the right. Note: the Column Manager is configured as a detector.
- On the Configure Communications page, click on Scan for Instruments. A window will populate with all the components that can be seen. Close this window.
- Assign the appropriate serial number from the pull-down list next to each detector.
- Click OK > Finish > Finish.
- In the inlet method, go to LC > Reset Communications.
- If it is open, close the ACQUITY Console, and then reopen it.
ADDITIONAL INFORMATION
id263023, comms, MLYNX, MLYNXV41, SUPMM