Source Gas On and Off tasks are missing from the MassLynx Shutdown Editor, and startup fails at source gas on task - WKB25826

Article number: 25826

SYMPTOMS

  • When MassLynx is first started, the Source Gas On (ESP) and Source Gas Off (ESP) options are missing from the Task drop-down list in the shutdown editor
  • This is a problem if you start MassLynx and then run a batch startup that specifies the Source Gas On task in the method
  • Unable to turn off API gas using shutdown method in MassLynx 
  • Startup or shutdown fails with error message:
    "Shutdown unable to complete execution of <startup/shutdown method name>
    Failed to turn ESP source gas on"

MassLynx must be restarted to clear the error.

Examples of releases that show the problem: MassLynx SCN 714, SCN 805, SCN 810, SCN 843, SCN 997, SCN 986, and SCN 1040

clipboard_ed8ea05d333c3c47ae775d325975752ed.png

ENVIRONMENT

  • 3100
  • ACQUITY TQD
  • Quattro Premier
  • Quattro micro
  • SQ Detector
  • SQ Detector 2
  • SYNAPT G2-S
  • Xevo G2-S QTof
  • Xevo TQ Absolute
  • Xevo TQ
  • Xevo TQD
  • Xevo TQ-S
  • Xevo TQ-S cronos
  • Xevo TQ-S micro
  • Xevo TQ-XS

CAUSE

Software defect CRI-3822 

The error "Failed to turn ESP source gas on" happens because the Source Gas On task is missing from the drop-down list but present in the method.

FIX or WORKAROUND

Fix:
This is fixed in SCN 1043 and subsequent software releases for all quadrupole instruments listed in the release notes.

It is also fixed for SYNAPT G2-S and Xevo G2-S QTof.

 

Workaround:
The Source Gas On and Off options appear in the drop-down list if you do one of the following:

  • Acquire a sample.

  • Toggle the ion mode on the tune page before running the Startup.

Note: The problem returns if MassLynx is restarted.
 

ADDITIONAL INFORMATION

 

id25826, A-SQ, A-SQ, MLYNX, MLYNXV41, Q3100MSD, QDA, QUATPREMXE, QUMICRO, QUPREMIER, RADIAN, SCN1040, SCN1043, SCN714, SCN805, SCN810, SCN843, SCN986, SCN997, SQD, SQD2, SUPMM, SYNG2SHD, SYNG2SHDML, SYNG2SIMS, SYNG2SMS, SYNG2SMSML, TQD, TQDIVD, XEVOG2SQTF, XEVOG2STOF, XEVOG2XSQT, XEVOTQ, XEVOTQA, XEVOTQA, XEVOTQD, XEVOTQDIVD, XEVOTQGC, XEVOTQIVD, XEVOTQS, XEVOTQSCRO, XEVOTQSIVD, XEVOTQSMIC, XEVOTQXS, XEVTQSMIVD, XEVTQXSIVD, ZQ2000, ZQ2000SFC, ZQ4000