How to set up a solvent delay method event in MassLynx - WKB23296
Article number: 23296
OBJECTIVE or GOAL
Direct the solvent flow to waste at the start of an injection, to avoid putting salts into the MS.
ENVIRONMENT
- MassLynx
- Xevo TQ-S
PROCEDURE
In MassLynx:
- Click MS method.
- Select Options > Method events.
- Click Enable.
- Confirm that the flow state in the Initial Settings is set to LC if making injection.
- Specify the following:
Time/mins: 0.01
Event: Solvent Delay
Action: Begin - Click Add to populate the table with the information from step 5.
- Specify:
Time/mins: 1.5 (or the desired end time)
Event: Solvent Delay
Action: End - Click Add to populate the table with the information from step 7.
- Click OK.
- Select File > Save or Save As to save the method.
- Make an injection.
Note: No data is collected while the solvent flow is directed to waste.