MassLynx loads an old calibration file - WKB22918

Article number: 22918

SYMPTOMS

  • The latest calibration file is open at the tune page, but the next time the tune page is open, an older calibration file is loaded.

ENVIRONMENT

  • MassLynx
  • TQD
  • Xevo TQ-S

CAUSE

An older calibration file is linked at the tune page.

FIX or WORKAROUND

  1. Launch the tune page.

  2. Select calibration, and then select set linked calibration.

  3. Browse to C:\MassLynx\Intellistart\Results\(the desired resolution mode) and select the latest calibration file. 

  4. Select link.

ADDITIONAL INFORMATION