Calibrations stuck at "creating calibration" in IntelliStart, do not complete - WKB229156
Article number: 229156
SYMPTOMS
- Calibrations stuck at "creating calibration" in IntelliStart, do not complete
ENVIRONMENT
- Xevo G2-XS
- MassLynx 4.2 SCN 971
CAUSE
Large number of calibration profiles present (>450)
FIX or WORKAROUND
- Close MassLynx.
- Delete unnecessary profiles from the C:MassLynx/Intellistart/Configuration/CalibrationProfiles folder.
- Run Stop Acquity processes.
- Open MassLynx and create new calibration.
ADDITIONAL INFORMATION
Advise users that they can reset and reuse old calibration profiles, instead of creating new ones each time.
id229156, SCN971, XEVOG2XSTF