Start a sample list in MassLynx, but the ACQUITY does not inject, and the MS goes straight to acquisition - WKB22465

Article number: 22465

SYMPTOMS

  • Start a sample set in MassLynx, but the ACQUITY does not inject
  • The MS goes straight to acquiring

ENVIRONMENT

  • MassLynx
  • LCMS acquisitions from the sample list

CAUSE

The events and triggering is not turned on.

FIX or WORKAROUND

  1. Go to the inlet editor and select the Tools drop-down menu.
  2. Select Instrument Configuration.
  3. Select Events and Triggering in the bottom-right corner.
  4. Select Events In 1 and next.
  5. For the pump, it is okay to leave it as Trigger by contact closure (even when not using a contact closure cable).
  6. Select Next, Finish, and Finish.

ADDITIONAL INFORMATION