Start a sample list in MassLynx, but the inlet does not inject, and the MS goes straight to acquisition - WKB22465
Article number: 22465
SYMPTOMS
- Start a sample set in MassLynx, but the ACQUITY does not inject
- The MS goes straight to acquiring
ENVIRONMENT
- MassLynx
- XEVOTQD
- SQ Detector
- ACQUITY TQD system
- TQ Detector
- SQ Detector 2
- Xevo TQ Absolute
- Xevo TQ MS
- Xevo TQD
- Xevo TQ-S
- Xevo TQ-S cronos
- Xevo TQ-S micro
- Xevo TQ-XS
CAUSE
The events and triggering is not turned on.
FIX or WORKAROUND
- Go to the inlet editor and select the Tools drop-down menu.
- Select Instrument Configuration.
- Select Events and Triggering in the bottom-right corner.
- Select Events In 1 and next.
- For the pump, it is okay to leave it as Trigger by contact closure (even when not using a contact closure cable).
- Select Next, Finish, and Finish.
ADDITIONAL INFORMATION
