MassLynx IntelliStart's Sample Tune and Develop Method uses the wrong tune page using the Xevo TQ-XS - WKB221644
Article number: 221644
SYMPTOMS
- When trying to do a Sample Tune and Develop method in IntelliStart, the system doesn't use the tune page specified
- Prior to running, can see optimized signals on the current tune page, but the IntelliStart results are poor
- When running, the tune page changes to an unrelated one in a different project that has different temperatures and voltage settings
- Renaming files without spaces and unsupported characters makes no difference
- Restarting the MS makes no difference
- Switching to a previous project makes no difference
ENVIRONMENT
- MassLynx 4.2 SCN 1012
- Xevo TQ-XS
- IntelliStart
CAUSE
Project corruption
FIX or WORKAROUND
- Make a new project with C:\MassLynx\Default.pro as the template.
- Create new files using appropriate naming scheme and ensure that files are saved to the new project location.
- Rerun the Sample Tune and Develop Method routine in IntelliStart.
ADDITIONAL INFORMATION