No results seen in TargetLynx - WKB208916

Article number: 208916

SYMPTOMS

  • After processing samples, the TargetLynx results file is empty; no calibration data is shown, summary table is empty
  • In Chromatogram, can see the peak of interest and can integrate it
  • In the TargetLynx method, the Acquisition function is correct for the compound
  • In the TargetLynx method, the Quantification trace information was manually entered and looks correct

ENVIRONMENT

  • MassLynx 4.2
  • TargetLynx

CAUSE

Incorrect syntax for Quantification trace

FIX or WORKAROUND

  1. Create a new compound in the Method Editor.
  2. Have the quantification trace field autopopulate with the correct syntax by right-clicking and dragging over the designated peak in the MassLynx Chromatogram.

ADDITIONAL INFORMATION