384-well plates cause axis errors for the ACQUITY Sample Manager - WKB206538
Article number: 206538
SYMPTOMS
- 384-well plates cause sample manager axis errors
ENVIRONMENT
- MassLynx
- ACQUITY Sample Manager
CAUSE
Plates are not properly defined.
FIX or WORKAROUND
- Take precise measurements of the dimensions of the plate to be used.
- Click the Inlet Method button in MassLynx.
- In the Inlet Method editor window, select the Acquity Sampler drop-down menu and choose Plate Generator.
- Scroll to the plate that most closely matches the plate definitions you want to use by clicking the Next Rack button.
- Modify the information to match the precise dimensions of the plate to be used.
- Save the rack when finished.