In TargetLynx, cannot see the internal standard concentrations calculated for analyte samples - WKB206610
Article number: 206610
SYMPTOMS
- Unable to get the concentrations of the internal standard compound calculated for analytes
- Can see a calibration curve for all compounds
- Can see concentrations calculated for the standard samples for all compounds
- The internal standard's calibration properties in the TargetLynx method show that the Concentration of Standard is set to Conc D
- The sample list has only the Conc D column populated for the standard samples
ENVIRONMENT
- TargetLynx
- MassLynx 4.1
CAUSE
Analyte samples have nothing in the Conc D column in the sample list.
FIX or WORKAROUND
Be sure to specify the internal standard concentration in both standard and analyte samples if you are using a specific sample list column for the Concentration of Standard in your TargetLynx method.