No TargetLynx results when processing data using a previously saved calibration curve file (.cbd) - WKB206256

Article number: 206256

SYMPTOMS

  • Processed a set to standards with TargetLynx to create a calibration curve

  • Exported the calibration curve information to a calibration curve file (.cdb)

  • Processed a set of samples, selecting the curve file previously exported  

  • No results appear in the TargetLynx file

ENVIRONMENT

  • MassLynx 
  • TargetLynx

CAUSE

TargetLynx Quantify Method and TargetLynx Quantify Curve are not from the same file, or a calibration curve file is exported from another results file.

 

FIX or WORKAROUND

The TargetLynx Method used to generate the curve file must be the same TargetLynx Method used to process the subsequent sample sets.