"Cannot import files of this type" error when importing CSV worklist into MassLynx - WKB20543
Article number: 20543
SYMPTOMS
- The MassLynx user is attempting to import a worklist with a CSV file type into MassLynx.
- They choose File > Import Worksheet and select the required file in the browser.
- The file is not imported and the following error message appears: "Cannot import files of this type."
ENVIRONMENT
MassLynx 4.1 and 4.2, all releases with LIMS interface.
CAUSE
In the file browser, the "Files of type" was set to "All Files (*.*)".
MassLynx was upgraded, so the"'Files of type" was set back to the default.
FIX or WORKAROUND
Change the "Files of type" to "Comma Delimited (*.CSV;*TXT)".
ADDITIONAL INFORMATION
LIMS interface converts the received HL7 worklist messages into CSV worklist files that are compatible with MassLynx.