MS method editor error on Xevo TQ-XS: Autodwell time is different from the last saved dwell time. Press yes to load the calculated dwell time - WKB199244
Article number: 199244
SYMPTOMS
- Error message when saving a MassLynx MS method editor or opening existing method:
"Autodwell time is different from the last saved dwell time. Press yes to load the calculated dwell time or press no to retain the saved dwell time"

- Happens with newly written and saved method on reopening
- Pressing yes changes the values, but the system does not save the changes
- Error can happen when changing a value in the method that should not affect the cycle time (e.g., changing a Function time window but not so much that it overlaps other functions)
- MassLynx 4.2 SCN 1007
ENVIRONMENT
- MassLynx
- Xevo TQ-XS
CAUSE
Software bug
FIX or WORKAROUND
Fixed in SCN 1035 and subsequent software
Workaround: Change the function times slightly until the error does not occur.
ADDITIONAL INFORMATION
Was CRI-3433