For an eλ PDA in MassLynx, cannot set a wavelength of more than 500 nm - WKB19881
SYMPTOMS
- Cannot enter a wavelength of more than 500 nm, even though it is designed to go to 800 nm
ENVIRONMENT
- ACQUITY UPLC PDA eλ Detector
- MassLynx
CAUSE
The detector was not configured correctly in the inlet editor.
If you do not "scan for instruments" and select the correct instrument during setup, MassLynx uses the default instrument type (maximum wavelength 500 nm).
FIX or WORKAROUND
- Deconfigure the detector.
- Close the inlet editor.
- Reopen the inlet editor.
- Go to Tools > Instrument Configuration > Configure, and select the ACQUITY PDA.
- On the next page, click "Scan for instruments". Find your PDA and select it.
- Doing this ensures that the correct instrument settings are loaded.
- If you do not do the scan, the system "assumes" that the PDA is the default detector with a range to 500 nm.
- Step through the end of the Configuration Wizard hitting Finish then Finish again.
- The inlet method will open with a method name Untitled. Do a File > Save As and save the new method for this configuration.
Note: The Default inlet method located in the Default.pro\acqudb folder was created for the original PDA detector with the range of 200 - 500 nm. Thus, you should not use this method to create new methods for the e-lambda PDA. Use the method file saved in Step # 7 to create any subsequent methods.
ADDITIONAL INFORMATION
The Instrument Control Software (ICS) is the same for ACQUITY UPLC PDA and ACQUITY UPLC PDA eλ detectors. MassLynx must recognize the detector as the extended-wavelength version to configure it correctly.
Methods created with a prior configuration will still not allow the wavelengths to go to the extended wavelengths. Methods will have to be recreated.
