Databridge crashes when converting MassLynx data for a Xevo TQD - WKB19904

Article number: 19904

SYMPTOMS

  • Databridge crashes when converting MassLynx data to either NetCDF or ASCI
  • The conversion fails

ENVIRONMENT

  • MassLynx 4.1 SCN 940 or SCN 962, or MassLynx 4.2 SCN 984 (i.e., SCNs used for control of the Xevo TQD)

CAUSE

Software bug

FIX or WORKAROUND

Fixed in SCN 985 and later

ADDITIONAL INFORMATION

If PC cannot be upgraded, a standalone PC can be used for data conversion.