What is the mass window about the Parent mass in an MRM method for a Xevo TQD in MassLynx? - WKB198408

Article number: 198408

ENVIRONMENT

  • Xevo TQD
  • MassLynx
  • Xevo TQ-S micro
  • IntelliStart

ANSWER

The mass window about the Parent mass in an MRM method is determined by the Span setting in the method. The default Span is 0, which is a +/- 0.05 amu window.

You can alter the span in the MS Method Editor by adjusting the span to 1 = +/- 0.5. Higher settings may include unwanted peaks.

ADDITIONAL INFORMATION

IntelliStart scans for the mass, selects ions, and then creates an MS Method with a default span of 0 (+/- 0.05 amu).

 