"Server Busy" error occurs when starting MassLynx - WKB197972

Article number: 197972

SYMPTOMS

  • A "Server Busy" error occurs when starting MassLynx

ENVIRONMENT

  • MassLynx 4.1
  • MassLynx 4.2
  • Windows 10

CAUSE

MassLynx cannot find the inlet systems.

FIX or WORKAROUND

  1. Exit MassLynx.

  2. Double-click StopAcquityProcesses.cmd on the desktop.

If you cannot find it on the desktop, it is located in C:\Program Files (x86)\Waters Instruments\Bin\.

  1. Turn off the main power for all inlet systems.

  2. Turn on the main power for all inlet systems.

  3. Launch MassLynx.

ADDITIONAL INFORMATION