MassLynx system does not inject after PM - WKB19776
Article number: 19776
SYMPTOMS
- Inlet does not inject
ENVIRONMENT
- MassLynx with any mass spectrometer
CAUSE
Triggering was disabled to perform the PM.
FIX or WORKAROUND
- Go to the inlet method and select Tools > Instrument Configuration > Events and Triggering. Select the "Event In 1" check box.
- Reboot the system: first the PC, then the inlet, then the Mass Spec (EPC).
ADDITIONAL INFORMATION
id19776, MLYNX, MLYNXV41, SUPMM