Skip to main content
Waters

In MassLynx, the MS acquires but no injection is seen - WKB197766

Article number: 197766

SYMPTOMS

  • New sample acquisition starts and the MS says it's acquiring, but there's no injection

ENVIRONMENT

  • MassLynx
  • ACQUITY UPLC
  • Xevo G2-S

CAUSE

The wrong syntax was used to define the position of the sample vial with the current bed layout.

FIX or WORKAROUND

  1. On the sample list, go to the bottle column and select the sample row of interest.
  2. Right-click and choose "Autosampler Bed Layout..." from the menu that appears. (In this example, the first row is selected and Autosampler Bed Layout is the last choice in the menu.)

Samplelist.png

3. In the AutoSampler Bed Layout screen, select the plate of interest. (In the example below, plate 1 is selected.)

autosampler bed layout.png

4. Right-click to select the well from this plate you want to use. (The well indicator changes from black to green.) 

5. In the menu that appears, choose the option "Replace".

autosampler bed layout add.png

6. Note that the Bottle column updates with a syntax; this is the correct syntax for the current set of plates defined in the bed layout. (In this example, note how the bottle position updates to 1:6.)

samplelist synatx updated.png

ADDITIONAL INFORMATION

 

id197766, MLYNX, MLYNXV41, SUPMM, XEVOG2SQTF, XEVOG2STOF

Not able to find a solution? Click here to request help.