No communications with MS; error "Could not connect to the Waters DHCP Server" - WKB19498
Article number: 19498
SYMPTOMS
- Installing SCN 849 on a new P510 PC
- Trying to run the Waters DHCP Server Configuration file in C:\Program Files (x86)\Waters Instruments
- Waters DHCP Server Configuration error box appears:
"Could not connect to the Waters DHCP Server
Windows error code:
The operation completed successfully"

- No communications to mass spectrometer
- Cannot connect to DHCP Server; "Cannot connect at this time" message appears
- Error in inlet editor: "Waters Generic Instrument COM Server - Detector Method get configuration failed with error: 80004005 unspecified error"
- PC and EPC restarted with no change
ENVIRONMENT
- MassLynx SCN 849 with compulsory patch SCN 896
- P510 PC
- Windows 7
- TQ Detector (Waters ACQUITY TQD)
CAUSE
Waters DHCP Server Service (WDHCP Service) did not start.
FIX or WORKAROUND
Restart the Waters DHCP File Service according to How to restart the Waters DHCP Server - WKB45004.