"Inlet method not ready" error during trapping stage for ACQUITY UPLC 2D M-Class - WKB194985

Article number: 194985

SYMPTOMS

  • 2D M-Class analysis starts, but an "inlet method not ready" error appears in the MassLynx status window during the trapping stage
  • The LED status lights on each module are green
  • In the MS console, the system status of each module is good, without any error

ENVIRONMENT

  • ACQUITY UPLC M-Class
  • MassLynx

CAUSE

The trigger method for the detector is set as Trigger by Software.

FIX or WORKAROUND

Set the trigger method for the detector as Trigger by Contact Closure.

ADDITIONAL INFORMATION

 