"Inlet method not ready" error during trapping stage for ACQUITY UPLC 2D M-Class - WKB194985
Article number: 194985
SYMPTOMS
- 2D M-Class analysis starts, but an "inlet method not ready" error appears in the MassLynx status window during the trapping stage
- The LED status lights on each module are green
- In the MS console, the system status of each module is good, without any error
ENVIRONMENT
- ACQUITY UPLC M-Class
- MassLynx
CAUSE
The trigger method for the detector is set as Trigger by Software.
FIX or WORKAROUND
Set the trigger method for the detector as Trigger by Contact Closure.
ADDITIONAL INFORMATION