How do you create a plate file for a 96-well plate with a well volume that is not one of the defaults? - WKB191362

Article number: 191362

OBJECTIVE or GOAL

Create a plate file (*.plt) for a 96-well plate with a well volume that is not one of the defaults.

ENVIRONMENT

  • MassLynx

PROCEDURE

  1. In the MassLynx Inlet Method, go to the "autosample" menu (ACQUITY Sampler, 2767, etc.) and select Plate Generator.
  2. Locate a 96-well plate using the next rack button on the tool bar.
  3. Enter a new name in the Rack Name field specific to the number of wells and well volume.
  4. Go to Rack > Save Current rack to save this file with the new file name.
  5. From documentation that comes with the rack, enter the height of the rack into the Z field under Plate Size and the Depth of the well. Values are entered in 1/10 mm. These are the only two values that differ between 96-well plates of different volumes. All other dimensions are the same.
  6. Save the settings.

ADDITIONAL INFORMATION

This procedure applies to any plate which does not have default *.plt file in the Racks folder for Masslynx.