ASAP probe acquisition error: Waiting for inlet start. Please reset the inlet system - WKB1897
Article number: 1897
SYMPTOMS
- ASAP probe acquisition does not start
- Error message: "Waiting for inlet start. Please reset the inlet system."
ENVIRONMENT
- MassLynx 4.1
- ASAP probe
CAUSE
No LC is required/permitted when using an ASAP probe. The LC must be unconfigured, or the events and triggering must be turned off. The latter is the better option.
FIX or WORKAROUND
- Go to the inlet editor and select Tools > Instrument configuration.
- Select the Events and Triggering button at the bottom-right > Next.
- Clear the selection for Event In 1, and then select Next > Next > Finish > Finish.
- Now, when you run from the sample list, it should not prompt you for details on the LC. You still need to fill in an inlet method and a bottle, but it won't use them.
ADDITIONAL INFORMATION