Skip to main content
Waters

Cannot quantitate UV data in TargetLynx - WKB1875

Article number: 1875

SYMPTOMS

  • No calibration curve when the sample set is processed

ENVIRONMENT

  • MassLynx 4.1
  • TargetLynx

CAUSE

Chromatogram mass window in m/z increments are 0.1 instead of nm (1), for example

FIX or WORKAROUND

  1. Click Yes for Use absolute mass window.
  2. In User Defined Properties, adjust the chromatogram mass window to 1.000.
  3. Save the method.
  4. Reprocess your data.

ADDITIONAL INFORMATION

In User Defined Properties, ensure that "Use absolute mass window"" is marked Yes.

To access these parameters, right-click within the User Defined Properties tab to view additional parameters

clipboard_e812a80bda3a7e5f9d5938e706b8bd29d.png

id1875, MLYNX, MLYNXV41, SUPMM, TARLYNX41

Not able to find a solution? Click here to request help.