Skip to main content
Waters

Intermittent communication issues with MS system - WKB17267

Article number: 17267

SYMPTOMS

  • Communication problems with LC-MS system
  • Regular, intermittent loss off connection with MS and modules in ACQUITY stack

ENVIRONMENT

  • MassLynx
  • ACQUITY UPLC modules
  • Kaspersky antivirus software

CAUSE

Unsupported and incorrectly configured antivirus software interrupts communications.

FIX or WORKAROUND

  1. Disable the antivirus software to confirm that it is the cause of the problem.
  2. Reconfigure the antivirus software:
    Exclude all folders and processes related to MassLynx software and instrument control.
    Disable real-time updates during instrument acquisition.
    Disable active scanning during data acquisition.
    Disable Network Threat Protection.

ADDITIONAL INFORMATION

id17267, comms, MLYNX, MLYNXV41, SUPMM, UPBINARY

Not able to find a solution? Click here to request help.