A new method is not visible in the OALogin window - WKB1588

Article number: 1588

SYMPTOMS

  • Created new MS method
  • Created new inlet method
  • Created new MS tune method
  • Created new OpenLynx (*.olp) method listing the new MS, inlet, and MS tune methods
  • OALogin window did not show the new method as an option

ENVIRONMENT

  • MassLynx 4.1
  • OpenLynx/OALogin

CAUSE

Need to select the new OpenLynx (*.olp) method for the user's group.

FIX or WORKAROUND

In the OALogin window, an administrator needs to do the following:

  1. Select Admin > User Configuration.
  2. Click the Group tab in the User Management window.
  3. Highlight the group listing the user.
  4. Click Edit Group.
  5. Click the Methods tab.
  6. Highlight the new OpenLynx method in the Available pane on the left.
  7. Click Add.
  8. Click OK to close the Edit Group window.
  9. Click OK to close the User Management window.
  10. Click Login Samples.

ADDITIONAL INFORMATION