Internal Standard not quantitating properly in TargetLynx - WKB15037
Article number: 15037
SYMPTOMS
- The same amount of internal standard is added to the sample and the standards.
- The internal standard amounts are not calculated properly. Skewing the results for the samples.
ENVIRONMENT
- QuanLynx
- TargetLynx
CAUSE
Internal Standard concentration is not set to fixed, but Conc A, which is varied in the sample list.
FIX or WORKAROUND
- In the TargetLynx or QuanLynx method, go to the Calibration Properties for the internal standard compound and set the concentration of standard:Level to Fixed (not Conc A).
ADDITIONAL INFORMATION
id15037, TARLYNX41