Is object storage/cloud storage supported with MassLynx? - WKB14967
Article number: 14967
ENVIRONMENT
- MassLynx
- EMC object storage
- Cloud storage
ANSWER
Real-time acquisitions with MassLynx are not supported with object storage or cloud storage and should remain on the acquisition PC.
Acquisition to servers is supported for quad data for MassLynx but not virtual servers.
ADDITIONAL INFORMATION
Archived MassLynx projects could use object storage, but customers need to test their retrieval. Object storage is flat (no file structure), so data retrieval/processing would be problematic.