MassLynx OALogin and AutoPurify system uses default volume instead of specified volume if using multiple injections - WKB13154
Article number: 13154
SYMPTOMS
- ACQUITY QDa with MassLynx SCN 947 in OALogin or Open Access / OpenLynx mode for Prep Systems
- Logging in a single injection, the system runs the injection volume specified in the OpenAccess login details
- The system uses the volume specified in the OpenLynx method (*.olp), not that in the login details, for all injections when multiple samples or injections are specified
- The system uses the volume specified in the OpenLynx method, not that in the login details, for all injections when multiple low-volume injections are specified instead of one high-volume injection
- The system uses the volume specified in the Prep method (*.olp) for all injections when using multiple injections
ENVIRONMENT
- Autopurification system
- Prep System
- MassLynx SCN 947 with and without SCN 961
- ACQUITY QDa
- OpenLynx / OALogin
- Multiple injection volumes
CAUSE
Bug in the software - PFD 18838, CRI-2786
FIX or WORKAROUND
Fixed in SCN 987 and reassessed in SCN 1046 (due for release Q2 2024)
Workaround: Log in samples individually.
If multiple samples must be injected from the same vial, make the vial location a non-mandatory entry in the OpenLynx method. After logging in the first injection, use the same vial location for subsequent injections.