In MassLynx, MaxEnt 3 Processing Freezes - WKB13009

Article number: 13009

SYMPTOMS

  • After starting to process spectrum data with MaxEnt 3, the software freezes.
  • It worked with the same data the day before.
  • Using the default.pro, HFN1.raw data set behaves the same way.

ENVIRONMENT

  • MassLynx 4.1 SCN 932
  • Synapt G2-Si

CAUSE

MaxEnt 3 Advanced Parameters were set to the following:

  • No. of Ensemble Members = 0
  • Iterations per Ensemble Member = 0
  • Compress Data = disabled

FIX or WORKAROUND

  1. In the Spectrum window, go to the Process tab.
  2. Click on MaxEnt 3.
  3. Set parameters as desired, and then click Advanced:

ME3 1.png

4. On the MaxEnt 3 Advanced Parameters window, change the settings to recommended defaults of:

ME3 2.png

5. Click OK to exit the MaxEnt 3 Advanced Parameters window.

6. On the MaxEnt 3 window, click OK to begin processing.

ADDITIONAL INFORMATION

Note: Once changed, you cannot go back to entering 0s for the number of ensemble members or iterations as the software will prompt you to enter a value between 1 and 100 (see below):

ME3 3.png