How to set up auto addition in MassLynx - WKB13017
Article number: 13017
OBJECTIVE or GOAL
To co-inject an aliquot from one vial with a sample from another vial using the MassLynx with any Sample Manager Flow Through Needle (FTN).
ENVIRONMENT
- MassLynx 4.1
- ACQUITY Sample Manager-FTN
PROCEDURE
In the MassLynx sample list:
- Specify the first sample draw as usual using the Bottle and Inject Volume columns.
- Add the column Auto Additions to the sample list. To add the column:
- Right-click in the sample list
- Select Customize Display.
- Scroll to Auto Additions.
- Select.
- Click OK.
- Right-click in the sample list
- Specify the additional sample draws in the Auto Addition column using the syntax:
- vial position, volume; vial position, volume
- Note: It is possible to repeat this three times for a maximum of four sample draws in one injection.
- vial position, volume; vial position, volume
ADDITIONAL INFORMATION
This feature is available for the Sample Manager Flow Through Needle (SM-FTN), but not compatible with fixed loop such as Sample Manager SM-FL.