Random flat regions in MS chromatogram with no data - WKB113183

Article number: 113183

SYMPTOMS

  • Random loss of signal in MS data shows as flat patches in an MS or MRM chromatogram
  • Peak retention times may be much shorter than expected; some peaks show as eluting earlier
  • Sometimes runs do not proceed to the next injection/sample
  • Instrument has intermittent communication problems

ENVIRONMENT

  • MassLynx
  • Xevo TQ-S

CAUSE

Faulty EPC/PCB

FIX or WORKAROUND

Replace the EPC (700011854).

ADDITIONAL INFORMATION

 

id113183, comms, MLYNX, MLYNXV41, SUPMM, XEVOTQS, XEVOTQSIVD, XEVTQSMIVD