Random flat regions in MS chromatogram with no data - WKB113183
Article number: 113183
SYMPTOMS
- Random loss of signal in MS data shows as flat patches in an MS or MRM chromatogram
- Peak retention times may be much shorter than expected; some peaks show as eluting earlier
- Sometimes runs do not proceed to the next injection/sample
- Instrument has intermittent communication problems
ENVIRONMENT
- MassLynx
- Xevo TQ-S
CAUSE
Faulty EPC/PCB
FIX or WORKAROUND
Replace the EPC (700011854).
ADDITIONAL INFORMATION