"Failed to run Analytical Component Engine" error when launching MassLynx - WKB111389
Article number: 111389
SYMPTOMS
- "Failed to run Analytical Component Engine" error message when launching MassLynx
- Inlet method does not open when clicking on the lozenge under the Instrument shortcut
- Inlet method does not open when right-clicking on a cell in the Inlet Method column in the sample list and selecting Edit from the pop-up menu
ENVIRONMENT
- MassLynx
CAUSE
Some type of corruption in the links between the InletEditor.exe file and the main MassLynx user interface (Sample List and Shortcut menus)
FIX or WORKAROUND
- Open the Task Manager (right-click on the Task bar at the bottom of the desktop and select Task manager from the menu.
- On the Processes tab, locate the InletKernal and the InletEditor processes. Highlight each in turn and select the Stop Process button in the lower-right corner.
- Browse to the C:\Masslynx folder.
- Locate the InletKernel.exe file. Double-click on the file.
- Locate the InletEditor.exe file. Double-click on the file.
- From the InletEditor, go to File > Open and open the Default inlet method for the specific configuration.
- Create a new Inlet Method file using the Default inlet method as a template.
ADDITIONAL INFORMATION