Poor sensitivity with TQD - WKB101894
Article number: 101894
SYMPTOMS
- Poor sensitivity with TQD
- Checked with the Setup or SetupQP reagent by infusion, the sensitivity is 1/10 or less of the normal value with the “RF lens” set to 0. Increasing the “RF lens” value such as 2 will increase the sensitivity 10 times or more.
ENVIRONMENT
- TQD
CAUSE
Hexapol contamination
FIX or WORKAROUND
- Clean/replace the source block and hexapole and set the “RF lens” to 0, the sensitivity also increase.
ADDITIONAL INFORMATION
