ANSWER

? = Peak exceeds dead time correction limits/potential errors
! = Interference or poor mass spectral resolution i.e. lockmass not detected or MS resolution is out of range (see WKB108561 for more information)
¿ = ADC is overloaded
* = Lockmass

ADDITIONAL INFORMATION

In the Spectrum window, find these definitions under the help menu by searching for "peak flags". 

id10187, MALDIG2SI, MLYNX, MLYNXV41, SUPMM, SYNAPT, SYNAPTMS, SYNAPTXS, SYNG2HD, SYNG2HDMAL, SYNG2MS, SYNG2MSMAL, SYNG2SHD, SYNG2SHDML, SYNG2SIHD, SYNG2SIMS, SYNG2SMS, SYNG2SMSML, XEVOG2QTOF, XEVOG2SQTF, XEVOG2STOF, XEVOG2TOF, XEVOG2XSTF, XEVOQTOF