How to configure the inlet in MassLynx without the PDA - WKB10185
Article number: 10185
OBJECTIVE or GOAL
Acquire in MassLynx without the PDA.
ENVIRONMENT
- MassLynx 4.1, 4.2
- PDA
- LC-MS
PROCEDURE
- Go to the inlet method editor.
- Select Tools > Instrument configuration.
- Select Configure and Next.
- Select Next to keep the pump configuration.
- Select Next to keep the autosampler configuration.
- Select the PDA from the Configured Detectors.
- Select the left arrow to remove the PDA and select Next.
- Select Next, Next. and Finish to complete the setup.
- Let the system initialize.
ADDITIONAL INFORMATION
This procedure does not remove the PDA from the stack in the console. To do so, power-off the PDA, and then close and reopen the console.
Re-boot the PC and LC to reset the communication.