How to Create a Component Custom Field - Tip84
Article number: 69103
OBJECTIVE or GOAL
Welcome back to Get Empowered! In the last Empower tip-of-the-week post for Empower CDS Software, we learned how to create a Custom Sample Set Field. (Tip #83)
Creating Custom Fields in Empower:
This week, we will learn how to create a Component Custom Field (Part 3).
A Component Custom Field is a real number that is related to a component(s). These values are typically used in subsequent Peak Custom Fields (calculations), which we will cover in a future tip. The perfect example is a common one – Label Claim. (We will use this value in a future peak field to calculate percent of Label Claim.)
Let’s get started.
ENVIRONMENT
- Empower
PROCEDURE
- Launch the new Custom Field wizard. On the Data and Type Selection screen select Component and the data type defaults to Real.

- On the Numeric Parameters screen, in addition to width and precision, we have the ability to enter a minimum and maximum value for this field. This would prevent someone from entering a value outside of the accepted range.

- Where are these values entered? After loading a Sample Set method in Run Samples click the Amount tool.

- Once the Component Editor opens, you will see a tab for the Component Field we created.

- Click on the tab for Label Claim. In order to enter the Label Claim for the samples, click the SampleSet Type drop-down list and select Standards & Unknowns.

- Enter the Label Claim value for the component(s). If you enter a value outside of the range specified when we created the Custom Field, a message is displayed reminding you of the acceptable range.

- We can enter the information post run in the Alter Sample window by clicking the Amount tool.

- Click the Label_Claim tab, change the SampleSet Type to Standards & Samples and then enter the Label Claim value for the component(s).

It’s that easy!
ADDITIONAL INFORMATION
- You can report these values by adding the Component Amounts Group to the report method.
- This procedure can be followed using the QuickStart or Pro interface. When using the QuickStart interface and in the Sample Queue, click the Components tool to access the Component Editor.
