How to save an instrument method with a method set - Tip12
Article number: 50271
OBJECTIVE or GOAL
Save an instrument method within a method set. (To inject a sample or run a Sample Set Method, we need a Method Set that contains an Instrument Method.
ENVIRONMENT
- Empower
PROCEDURE
- From the Pro Interface, select Browse Projects.

- Select a project from the list and click OK.

- Select the Methods tab.

- Right-click in the table and select New Method > Instrument Method.

- If you have more than one chromatography system, select the one you want to create the new Instrument Method for and click OK.

- Specify the parameters for the new Instrument Method. From the File menu, select Save with Method Set.

- Specify a name for the new Instrument Method, type any pertinent comments, and then click Save.

- Empower prompts you to enter a name for the Method Set. Type the name and any pertinent comments, and then click Save.

- The new Instrument Method and Method Set now appear in the Methods table.

- You can also do this from the Run Samples window by clicking Edit under the Instrument Method field and following steps 6 through 9 above.

ADDITIONAL INFORMATION
You must log in to Empower with the Pro Interface to take advantage of this feature.
