System Policies: New Projects - Tip380
OBJECTIVE or GOAL
System Policies for New Projects
ENVIRONMENT
- Empower
- Empower Tip of the Week #380
PROCEDURE
System Policies for New Projects only affect new projects that are created. These do not affect existing projects.
STEP 1
Let’s look at the Default Full Audit Trail settings first (figure 1).
- Full Audit Trail Support – with this policy enabled, anytime someone creates a new project it will have a Project Audit Trail.
- Default Full Audit Trails Table – select whether a comment is required for Methods, Results, Samples and/or Deletion. For example, if you create or modify a Method, you may be required to enter a comment as to the reason why you performed this action.

STEP 2
The choices for comments are as follows (figure 2).
- Silent – no comment needed
- Unrestricted – enter text
- Restricted – pick from a list of pre-defined comments (more about this in steps 4-6).
NOTE: If Confirm Identity is selected, the User will need to enter their Empower password after the action is performed.

STEP 3
The selections below the table are as follows (figure 3).
- Don't allow user to change default Full Audit Trail Support setting – if this is enabled the User cannot change the Full Audit Trail Support selection made in Step 1, bullet 1.
- Don't allow user to change default 'Require User Comments On' settings – if this is enabled the User cannot change the selections made in Step 1, bullet 2.
- Don't allow user to copy from non-FAT projects into FAT projects – if this is enabled the User cannot copy data from a project without an Audit Trail into a project which has an Audit Trail.
- Allow shallow copies between FAT projects – if this is enabled only the latest version of a method is copied between projects which have Audit Trails.

STEP 4
In Step 2, pre-defined comments have been mentioned which are called Default Strings. To create a Default String, go to Configuration Manager and select Default Strings from the View menu (figure 4).

STEP 5
Select the category from the list for which you want to create a Default String. In this example, it is Processing Methods (figure 5).

STEP 6
Click ‘Add String’ to open a dialogue box into which you enter the text. Click ‘OK’ and this comment will now be available when you create or modify a Processing Method in the project (figure 6).

ADDITIONAL INFORMATION
N/A