OBJECTIVE or GOAL

Calculate Amount/Concentration, Calibration Curve fit types and review Quadratic Fit. 

ENVIRONMENT

  • Empower
  • Empower Tip of the Week #280

PROCEDURE

STEP 1
For a Quadratic Fit, Empower performs a least-squares fit of a quadratic polynomial to the calibration points. A minimum of five points is recommended (Figure 1).

Figure_1.png

STEP 2
The Quadratic Fit generates a Calibration Curve that is represented by the equation in Figure 2.

Figure_2.png

STEP 3
Calculating amount or concentration in the sample is determined by solving for x in the equation. Empower reports only positive values of X that are within the range of the Calibration Curve (Figure 3).

Figure_3.png

STEP 4
Calibration Curves can be accessed in various ways, for example:

  • Review – Bring a result into Review and click on the Calibration Curve tool.
  • Compare – Right-click on a curve from the Curves tab in the project and select Compare.
     

ADDITIONAL INFORMATION

This can be done with either the Pro or QuickStart interface.

id235238, EMP2LIC, EMP2OPT, EMP2SW, EMP3GC, EMP3LIC, EMP3OPT, EMP3SW, EMPGC, EMPGPC, EMPLIC, EMPOWER2, EMPOWER3, EMPSW, SUP