Skip to main content
Waters

Calculating Amount/Concentration - Calibration Curve Fit Types: Linear Fits - Tip279

Article number: 235191

OBJECTIVE or GOAL

Calculate Amount/Concentration, Calibration Curve fit types and review Linear Fits. 

ENVIRONMENT

  • Empower
  • Empower Tip of the Week #279

PROCEDURE

STEP 1
For a Linear Fit, Empower calculates the line that best fits the amount (or concentrations) versus the responses. The equation is Y = Bx + A, where B is the slope and A is the Y intercept (Figure 1).

Figure_1.png

STEP 2
For an Inverse Linear Fit, Empower performs a Linear Fit to the X and Y values of the calibration points. The X value used in this case is 1/X for each standard peak. The equation used is the same as that for Linear Fit (Figure 2). 

Figure_2.png

STEP 3
For a Log-Log Fit, Empower performs a Linear Fit to the X and Y values of the calibration points. The Y value of each point is the Log of the response of the peaks in the standard. The X value of each point is the Log of the amount or concentration of the peaks in the standard. The equation used is the same as that for Linear Fit (Figure 3). 

Figure_3.png

STEP 4
For a Power Curve Fit, Empower performs a Linear Fit of log Y versus log X, where Y is the response and X is the X value (Figure 4).

Figure_4.png

STEP 5
The plot and the equation of the curve are reported in terms of the X value and response (Figure 5).

Figure_5.png
 

ADDITIONAL INFORMATION

This can be done with either the Pro or QuickStart interface.

Not able to find a solution? Click here to request help.