Skip to main content
Waters

Exploring parameters within the Sample Set Method that can affect the calculation of amount or concentration - Tip268

Article number: 229116

OBJECTIVE or GOAL

This article explores parameters within the Sample Set Method that can affect the calculation of amount or concentration. The parameters we will explore are Purity, Sample Weight and Dilution. 

ENVIRONMENT

  • Empower
  • Empwer Tip of the Week #268

PROCEDURE

Purity
STEP 1
The Purity within the Component Editor is where you enter a purity factor for the standards. The amount entered on the ‘Value’ tab is multiplied by the ‘purity factor’ and those values will be used for the ‘x axis’ of the calibration curve (figure 1). 
Figure_1.png


STEP 2
Once the Sample Set has been processed, view the ‘x Values’ in the calibration curve point table. As an example, for ‘Standard 1’ the entered amount was ‘0.0400’ and the value used (once corrected for the purity value) is ‘0.0396’ (figure 2).
Figure_2.png

STEP 3
The calculated amount in the sample is ‘0.099’ compared to ‘0.100’ which we saw in Tip #267 (figure 3).
Figure_3.png

Sample Weight and Dilution
STEP 4

When working with standards, the amount entered in the Component Editor is multiplied by the ‘Sample Weight’ entered in the Sample Set Method and then divided by the ‘Dilution’, also entered in the Sample Set Method, and that final value is used for the ‘x axis’ of the calibration curve (figure 4).
Figure_4.png

STEP 5
Once the Sample Set is processed, view the ‘x values’ in the calibration curve point table. For example, for ‘Standard 1’ the entered amount was ‘0.0400’ and the value used (once corrected for Purity, Sample Weight and Dilution) is ‘0.0792’ (figure 5). 
Figure_5.png
STEP 6
The calculated amount in the sample is now ‘0.198’ (figure 6)
Figure_6.png

STEP 7
When working with samples the amount determined from the calibration curve is divided by the ‘Sample Weight’ entered in the Sample Set Method and then multiplied by the ‘Dilution’ entered in the Sample Set Method. That is the final value that Empower displays (figure 7). 
Figure_7.png

STEP 8
The calculated amount in the sample is now ‘19.824’ (figure 8).
Figure_8.png
 

ADDITIONAL INFORMATION

In last week’s tip, we discussed How Empower calculates amount or concentration, Tip #267.

If SampleWeight field name changes to anything else Amount will not calculate correctly. 

Final Note: This can be done with either the Pro or QuickStart interface.

Not able to find a solution? Click here to request help.