Skip to main content
Waters

Understanding the difference in Resolution when working with a Photodiode Array Detector - Tip250

Article number: 221918

OBJECTIVE or GOAL

In this week’s tip, we will answer a question from an Empower user. 

Question: In the instrument method for my ACQUITY PDA, I set the resolution to 1.2nm, however, when I look at the resolution in Spectrum Review, I see 0.6nm. Why the difference?

Let's find out. 

ENVIRONMENT

  • Empower
  • PDA Detector
  • Photodiode Array Detector
  • Empower Tip of the Week #250

PROCEDURE

STEP 1
1.2nm is the lowest selectable value for Resolution in the ACQUITY PDA Instrument Method. This is the optical resolution (figure 1).

clipboard_e9d10344d25f8d8c2d686e33e85a15ceb.png

STEP 2

The spectra tab within Spectrum Review displays a Resolution of 0.6nm. Additionally, when we look at the spectrum points tab, we see the distance between consecutive points on a spectrum is approximately 0.6nm. This is the digital resolution which in the case of the ACQUITY PDA is half of the optical resolution selected (figure 2). 

clipboard_e88015565211f03583accc65f99b32fc6.png

STEP 3
How is this calculated? The formula used is as follows (figure 3): 

clipboard_ef8fbf92c0c8487252ceb82720e21d8ce.png

The result is rounded to 1 decimal.

STEP 4
Where do I find these values? If we look at the Spectra tab in the Spectrum Review, we see the maximum and minimum wavelengths for the spectrum (figure 4).

clipboard_ef156daeaefcd53235897904e7df7afcc.png

STEP 5

If we look at the Spectrum Points tab and scroll to the bottom of the table, we see the number of points for the spectrum, in this case, 310 (figure 5). 

clipboard_e7726378b64609f8f15b1caa1e8fa31ab.png

STEP 6

Substituting in the numbers we get 0.6 (figure 6).

clipboard_e397247b4dfe1224a342f5853a8df40b1.png

STEP 7

Now let’s compare the above to the 2998 PDA. 1.2nm is the lowest selectable value for Resolution in the Instrument Method. This is the optical resolution (Figure 7). 

clipboard_e74a44043c15282679659e454a4e8b10d.png

STEP 8

Looking at the Spectra tab in Spectrum Review we see the maximum and minimum wavelengths for the spectrum (figure 8). 

clipboard_eddd100ad9e0a2193d569ba19b34fea67.png

STEP 9

If we look at the Spectrum Points tab and scroll to the bottom of the table, we see the number of points for the spectrum, in this case, 161 (figure 9). 

clipboard_e7a95b30c02d5c39044877161fc01ea4d.png

STEP 10

Substituting in the numbers we get 1.2 (figure 10). 

clipboard_ef1805468c9f214aef1011dd073d6c78e.png

It is that easy!

 

ADDITIONAL INFORMATION

Final Note: This can be done with either the Pro or QuickStart interface. 

Not able to find a solution? Click here to request help.