Processing method picks up lots of peaks in Empower - WKB99966

Article number: 99966

SYMPTOMS

  • Processing method picks up lots of peaks in Empower

ENVIRONMENT

  • Empower 3

CAUSE

The processing method integration parameters are not set properly.

FIX or WORKAROUND

  1. Add a function Inhibit Integration with the time range in which you don't want to integrate.
  2. Refine the processing method with higher threshold and peak width.

ADDITIONAL INFORMATION

Note: unwanted peaks or large number of peaks on a chromatogram may leads to a slow performance or cause trouble to review and process sometimes.hence design the processing method as per your exact requirement.

 