For GC FID or FPD detector, Lit offset value is not available in the ICF instrument method - WKB99975
Article number: 99975
SYMPTOMS
- Lit offset value is not available in ICF instrument method for GC FID or FPD detector
ENVIRONMENT
- Agilent Instrument Control Framework v3.0 and above
- Empower 3 Enterprise
CAUSE
N/A
FIX or WORKAROUND
For FID or FPD detector in ICF, the Lit offset must be set on the GC front panel. Ensure that the hydrogen/air ratio is 8–12% to keep the flame lit. No setting is available in the ICF instrument method.
ADDITIONAL INFORMATION
After setting the Lit Offset value on the GC front panel, do the reconfiguration and create a new instrument method.
