Skip to main content
Waters

Empower 3: Is it possible to quantify using different processing methods for Standard and Unknown? - WKB99917

Article number: 99917

ENVIRONMENT

  • Empower 3

ANSWER

No, this is not possible.
Since the calibration curve is saved along with the processing method, it is necessary to create the calibration curve and perform quantification using the same processing method.


In addition, the system name and channel name must be the same for the Standard used for the calibration curve and the Unknown for the quantification.
If they are different, you cannot use the calibration curve even if you use the same processing method.

ADDITIONAL INFORMATION

 

Not able to find a solution? Click here to request help.