When running an acquisition on an Agilent 7890/6890 GC with an Agilent Headspace sampler, the system doesn't plot the channel data - WKB98429
Article number: 98429
SYMPTOMS
- Agilent GC 7890/6890 with headspace does not plot data for Agilent 2.6 Update 2 Driver or Agilent ICF v3.3
ENVIRONMENT
- Empower 3 Enterprise
- Agilent Instrument Control Framework
- Agilent 7890 GC
- Agilent 6890 GC
- Agilent 6850 GC
- Agilent 7697A Headspace
- Agilent G1888 Headspace
CAUSE
This fix comes from Agilent and Agilent Defect IDs 293090 and 293241.
FIX or WORKAROUND
- For Agilent GC 7890 with HS configuration, customer can use procedure below as a workaround:
The user should not enter the same source for both signals (irrespective of save enabled/disabled).
2. For 6890/6850 with HS configuration, use the Agilent ICF support layer v3.2 with Agilent ICF A.02.05, which incorporates GC Driver version B.01.03.a, on the acquisition client that controls the 6890GC/6850GC with the Agilent Headspace Sampler.