In Empower, the WFM-A instrument method freezes (becomes unresponsive) when the "Collection Event Table/Simulation" button is selected - WKB97799
Article number: 97799
SYMPTOMS
- The WFM-A instrument method freezes (becomes unresponsive) when the "Collection Event Table/Simulation" button is selected
- Cannot close or modify the instrument method after clicking "Collection Event Table/Simulation"
ENVIRONMENT
- Empower 3 Network
- Waters Fraction Manager - Analytical (WFM-A)
- Application server RDS (Remote Desktop Server) (only Citrix is officially supported with Empower 3)
CAUSE
Users sometimes use two screens, sometimes only one screen. If the window "Collection Event Table/Simulation" is moved to the second screen, it does not appear when only one screen is used.
FIX or WORKAROUND
Do not move the window "Collection Event Table/Simulation" to the second screen.
ADDITIONAL INFORMATION