In Empower, the WFM-A instrument method freezes (becomes unresponsive) when the "Collection Event Table/Simulation" button is selected - WKB97799

Article number: 97799

SYMPTOMS

  • The WFM-A instrument method freezes (becomes unresponsive) when the "Collection Event Table/Simulation" button is selected
  • Cannot close or modify the instrument method after clicking "Collection Event Table/Simulation"

ENVIRONMENT

  • Empower 3 Network
  • Waters Fraction Manager - Analytical (WFM-A)
  • Application server RDS (Remote Desktop Server) (only Citrix is officially supported with Empower 3)

CAUSE

Users sometimes use two screens, sometimes only one screen. If the window "Collection Event Table/Simulation" is moved to the second screen, it does not appear when only one screen is used.

clipboard_e3ef2121eb0de34d223a160eddb9454a2.png

FIX or WORKAROUND

Do not move the window "Collection Event Table/Simulation" to the second screen.

ADDITIONAL INFORMATION

 